5.2.2 Noise Cancel Procedure
When the detector gives a repetitive warble without the coil being
moved, perform the following electromagnetic Noise Cancel procedure:
CAUTION
Keep the detector very still; do not handle the detector during this
procedure. It can take some time and any movement can upset the
process, as will any metallic objects being moved in the vicinity of the
coil. Also ensure that there are no metallic objects near the coil during
this procedure.
a. Place the detector coil flat on the ground and press the blue Noise
Cancel button.
b. The detector will take 70 seconds to scan the locality and adjust the
detector to a quiet setting. During this time, a continuous, steady
tone can be heard.
c. The detector sounds four beeps once scanning is completed.
d. When this procedure is completed it may again be necessary to
follow the Audio Reset and Ground Balance procedures to
readjust the audio threshold and Ground Balance.
e. If all other settings have been established, test the detector with the
test piece (as described in section 5.4). If the detector test is
satisfactory it is now ready for use.
CAUTION
The F1A4 cannot detect targets while the Noise Cancel procedure is
being performed.